Digital Ohm Meter
View Large
Product Name : Digital Ohm Meter
Product Origin : China Taiwan
Product Category : Chitai
More Products : Click here!
Contact Details : 運揚有限公司 WANCO LIMITED
Product Code : 5601
Product Details































Digital Ohm Meter 5601
FEATURES
‧Maximum Display of 33000

‧Power Supply AC90V~260V, 50/60Hz

‧Wide Measurement range: From 10μΩ resolution to 30MΩ full scale

‧High accuracy ±0.05%

‧Auto/Manual Function

‧RS-232 Interface standard

‧Measurement Speed 10 samples/sec.

‧HOLD, REL Function
5601 Specifications RANGES
Maximum Display: 33000

Sampling Rate: 10 samples/sec.

Display: 6-digit, 7-segment LED 14.2mm High

Over Indication: "00000" flash

Range Selection: Automatic and Manual

Overload Protection: AC330Vrms

Power Supply: AC90V~264V, 50/60Hz<15VA

Operating temperature: 0~50℃ (32~122℉), RH≦80%

RS-232: Baud rate 300, 2400, 9600 No parity, One stop bit.

Dimension: 248mmL×93mmH×295mmD





































































Range


Text Range


Resolution


Text Current


Accuracy


300mΩ


10uΩ~330.00mΩ


10uΩ


DC 100mA


±0.05%

±20 digit




10uΩ~3.3000Ω


10uΩ


DC 100mA


±0.02%

±10 digit








30Ω


1mΩ~33.000Ω


1mΩ


DC 100mA


300Ω


10mΩ~330.00Ω


10mΩ


DC 10mA


3KΩ


100mΩ~3.3000KΩ


100mΩ


DC 1mA


30KΩ


1Ω~33.000KΩ




DC 1mA


300KΩ


10Ω~330.00kΩ


10Ω


DC 10uA


3MΩ


100Ω~3.3000MΩ


100Ω


DC 1uA


30MΩ


1KΩ~33.000MΩ


1KΩ


DC 100mA


±0.05%

±20 digit




SECTION OPERATION
1. Range selection

The 5601digital ohm meter has two range modes: Manual and Auto range. Measurement range can be selected using either auto range (pressing the-Auto button) or manual range (pressing the ▲or▼button).

In autorange, the ohm meter higher range when the reading is at full scale 30000 counts, and lower range when the rending below 9% of full scale 2700 counts, while the instrument changes range. The auto range up to the hightest range 30M ohms,

down to the 300m ohms.

Resistance measurements from 10μ ohms to 30M ohms in 9 range: 300m ohms, 3 ohms, 30 ohms, 300 ohms, 3K ohms, 30K ohms, 300K ohms,

3M ohms, 30M ohms.



2. HOLD Modifier


The 5601 is operating in a display hold mode, the Hold modifier allows you to take a measurement and "hold" that measurement on the display.



3. Four-Wire Measurement


The 5601 makes 4 wire ohms measurements as known in Figure 1.

The source HI and LO leads apply a known, internal current source to the unknown resistance, the voltage drop across the unknown resistance.